Influence of tip wear of piezoelectric ultrasonic scalers on root surface roughness at different working parameters. A profilometric and atomic force microscopy study


Creative Commons License

Arabacı T., Cicek Y., Dilsiz A., Erdogan I. Y., Kose O., Kizildag A.

INTERNATIONAL JOURNAL OF DENTAL HYGIENE, cilt.11, sa.1, ss.69-74, 2013 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 11 Sayı: 1
  • Basım Tarihi: 2013
  • Doi Numarası: 10.1111/idh.12003
  • Dergi Adı: INTERNATIONAL JOURNAL OF DENTAL HYGIENE
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.69-74
  • Anahtar Kelimeler: atomic force microscopy, surface roughness, tip wear, ultrasonic scaler, DISPLACEMENT AMPLITUDE, SCANNING-ELECTRON
  • Recep Tayyip Erdoğan Üniversitesi Adresli: Hayır

Özet

Roughness on tooth surfaces is reported to facilitate the reestablishment of microbial dental plaque. Hence, the main goal of dental scaling is to remove bacterial plaque and obtain smoother tooth surfaces. This study was aimed to assess the influence of tip wear of ultrasonic scaler inserts on root surface roughness at different working parameters. Twenty piezoelectric ultrasonic scaler inserts (10 worn/10 new) were selected to examine the erosion ratio (ER) on the scaler tips and to assess the influence of tip wear on root surface roughness. Erosion on the tip surfaces was evaluated under atomic force microscopy (AFM). Root samples were prepared and instrumented by new (Group I) and worn (Group II) inserts at different working parameters. Roughness change (Rc) on root surfaces after instrumentation was examined under profilometer and compared between and within the groups. Statistically significant differences were found between the mean ERs of new and worn tips (P < 0.01). The results of this study showed that tip angulation and instrument power strongly influenced the Rc values on instrumented samples (P < 0.05). It was also revealed that tip wear influenced the Rc values on root surfaces especially at 45 degrees tip angulation (P < 0.05). Therefore, tip wear should also be considered as much as the other parameters to minimize the surface roughness during ultrasonic treatment.

Abstract: Roughness on tooth surfaces is reported to facilitate the
reestablishment of microbial dental plaque. Hence, the main goal of
dental scaling is to remove bacterial plaque and obtain smoother
tooth surfaces. This study was aimed to assess the influence of tip
wear of ultrasonic scaler inserts on root surface roughness at different
working parameters. Twenty piezoelectric ultrasonic scaler inserts
(10 worn/10 new) were selected to examine the erosion ratio (ER) on
the scaler tips and to assess the influence of tip wear on root surface
roughness. Erosion on the tip surfaces was evaluated under atomic
force microscopy (AFM). Root samples were prepared and
instrumented by new (Group I) and worn (Group II) inserts at different
working parameters. Roughness change (Rc) on root surfaces after
instrumentation was examined under profilometer and compared
between and within the groups. Statistically significant differences
were found between the mean ERs of new and worn tips (P < 0.01).
The results of this study showed that tip angulation and instrument
power strongly influenced the Rc values on instrumented samples
(P < 0.05). It was also revealed that tip wear influenced the Rc values
on root surfaces especially at 45° tip angulation (P < 0.05). Therefore,
tip wear should also be considered as much as the other parameters
to minimize the surface roughness during ultrasonic treatment.
Key words: atomic force microscopy; surface roughness; tip wear;
ultrasonic scaler